180
Specifications
Standard Models
General Specifications:
AC power source/power requirements AC adapter, 90 V AC (min.) to 264 V AC (max.)
Frequency 48 60 Hz
Allowable ambient operating temperature 0 ... +40 °C (273 ... 313 K, 32 °F ... 104 °F)
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update 0.1 0.4
(depends on the filter level selected)
Power consumption 25 VA
Hours of operation with fully charged
YRB 06 Z external battery pack, approx. 14 h
Selectable weight units Grams, kilograms, carats, pounds, ounces, Troy ounces, Hong Kong taels, Singapore taels,
Taiwanese taels, grains, pennyweights, milligrams, parts per pound, Chinese taels, mommes,
Austrian carats, Tola, Baht and Mesghal
Selectable application programs Mass unit conversion, counting, weighing in percent, animal weighing, calculation,
density determination, over/under checkweighing, time-controlled functions, totalizing, statistics,
2nd tare memory, identifiers, product data memories
Built-in interface RS-232C
Format: 7-bit ASCII, 1 start bit,
1 or 2 stop bits
Parity: odd, even or space
Transmission rates: 150 to 19,200 baud
Handshake: Software or hardware
Specifications of the Individual Models:
Model FC06BBE-SX FC6CCE-HX FC2CCE-SX FC12CCE-SX FC6CCE-SX
Readability g 0.001 0.01 0.01 0.1 0.1
Weighing capacity g 620 6,200 2,200 12,000 6,200
Max. capacity kg 3 25 10 50 50
Tare range (subtractive) g 620 6,200 2,200 12,000 6,200
Electronically compensated preload g 93 110 1,200 1,240
(without restricting weighing range)
Max. preload when starting g 110 5,200 1,300 10,000 6,400
calibration/adjustment
(scale must be zeroed)
Repeatability (standard deviation) ±g 0.001 0.01 0.01 0.05 0.05
Linearity ±g 0.002 0.02 0.02 0.2 0.1
Sensitivity drift within +10 ... +30 °C ±/K 2 · 10
-6
2 · 10
-6
2 · 10
-6
4 · 10
-6
4 · 10
-6
Response time (average) s 1.5 1.5 1.5 1 1
External calibration weight g 500 (E2) 5,000 (E2) 2,000 (F1) 5,000 (F1) 5,000 (F2)
(of at least accuracy class...)
Load plate mm Ø 130 218 x 200 218 x 200 218 x 200 218 x 200
Dimensions (W x D x H) mm 240x294x86 240x294x86 240x294x86 240x294x86 240x294x86
Net weight, approx. kg 7 8.4 7.3 7.3 7.3
Dust and water protection rating
according to EN 60529 IP54

Table of Contents

Other manuals for Sartorius Factory Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Factory Series and is the answer not in the manual?

Sartorius Factory Series Specifications

General IconGeneral
BrandSartorius
ModelFactory Series
CategoryScales
LanguageEnglish

Related product manuals