EasyManua.ls Logo

Sartorius Factory Series - Weighing in Percent

Sartorius Factory Series
213 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
85
Operating the Scale
Weighing in Percent B
Purpose
This application program allows
you to obtain weight readouts
in percent which are in proportion to
a reference weight. Alternatively,
you can have the value displayed
as a difference in percent between
the weight on the scale and
the reference weight, or as a
special ratio 1 or ratio 2.
You can use this application program
in combination with a program
chosen from Application 2
(checkweighing, time-controlled
functions) and one from Application 3
(totalizing, formulation, statistics).
Available Features
– Reference percentage “pRef”
loaded from long-term memory
when you turn on the scale
– Optional scale configuration in
Setup for automatically initializing
this application and loading the
most recent reference percentage
“pRef” entered with reference
weight “Wxx%” when you turn on
the scale
– Value displayed as:
– Residual quantity (portion)
– Difference (deviation)
– Ratio1
– Ratio2
depending on the Setup menu
code selected.
– Reference percentage “nRef”
entered manually
– Store the current weight as the
reference percentage weight
“Wxx%” for initializing the
weighing-in-percent application
program
– Reference weight “Wxx%”
entered manually
– Storage parameter (rounding-off
factor) for storing the reference
weight “W100%” in percentage
calculation can be configured
– Configuration of decimal places
displayed with a percentage
– Optional configuration for having
the reference weight “Wxx%”
and reference percentage
automatically output via the data
interface port after initialization of
the weighing-in-percent program
(print application parameters)
– Toggle the display between
percentage and weight readout
by pressing the
Weigh. soft key
– Toggle between the
weighing-in-percent program
and other applications (e.g.,
checkweighing) by pressing D
Factory Settings
Storage parameter:
Display accuracy
Digits displayed with percentage:
2 digits
Display calculated value:
Residue
Soft Key Functions
pRef Store value input as
reference percentage
Wxx% Store input value as
reference sample weight
Perc. Toggle to the Weighing-
in-percent application
New Store next value
Weigh. Toggle to the Weighing
mode
Start Store the current weight
value for the preset refe-
rence sample quantity