80
Application Programs
Soft Key Functions
Start Start application program
Weigh. Toggle to basic
weighing functions
Auto-Start Application When the
Power Goes On
In the Setup menu, you can select
whether the application that is active
before you turn off the power will
automatically start when the power is
turned on again (Setup: Application
parameters: Auto-start app. when
power goes on: On).
Using Verified Scales as Legal
Measuring Instruments in the EU*:
All application programs can be
selected on scales verified for use
in legal metrology.
Non-metric values can be
indicated as follows:
– Percent = %
– Piece counting (Counting) = pcs
– Computed value = o, G
Toggle between Two
Weight Units
Purpose
With this application program you
can switch the display of a weight
value back and forth between two
weight units by pressing a soft key.
You can use this application program
in combination with a program
chosen from Application 2 (check-
weighing, time-controlled functions)
and one from Application 3
(totalizing, formulation, statistics).
Available Features
– Toggling the displayed weight
– Other features as for the basic
weighing function
Factory Settings
Weight unit 1:
Grams /g
Weight unit 2: Kilograms /Kg
* including the Signatories
of the Agreement on the European
Economic Area
Preparation
Scales used as legal measuring instruments: grams and kilograms are the
only weight units available
Standard scales: The following weight units are available in both ranges:
Line for
Display/ metrological
Unit Conversion factor Printout data
Grams 1.00000000000 g g
Kilograms 0.00100000000 kg kg
Carats 5.00000000000 ct ct
Pounds 0.00220462260 lb lb
Ounces 0.03527396200 oz oz
Troy ounces 0.03215074700 ozt ozt
Hong Kong taels 0.02671725000 tlh tlh
Singapore taels 0.02645544638 tls tls
Taiwanese taels 0.02666666000 tlt tlt
Grains 15.43235835000 GN GN
Pennyweights 0.64301493100 dwt dwt
Milligrams 1000.00000000000 mg mg
Parts per pound 1.12876677120 /lb lb
Chinese taels 0.02645547175 tlc tlc
Mommes 0.26670000000 mom M
Austrian carats 5.00000000000 K K
Tola 0.08573333810 tol tol
Baht 0.06578947437 bat bat
Mesghal 0.21700000000 MS MS
§ Turn on the scale: Press e
> Sartorius logo is displayed
§ Select the “Toggle weight units” program in the Setup menu: Press M
FC Models:
§ Select the Application menu: App soft key
§ Select Application 1: O soft key
FCA Models:
§ Select Application parameters: press the q soft key 2 x,
then the O soft key once
§ Application 1 (basic settings): press the O soft key
§ Select Toggle wt. units: Q or q soft key (repeatedly)
§ Confirm Toggle wt. units: O soft key
§ Select and confirm:
–
Weight unit 1: see above
–
Weight unit 2: see above
see also the “Application Menu (Overview)” in the chapter entitled
“Configuring the Scale”
§ Save settings and exit the Setup menu: Press the oo soft key