EasyManua.ls Logo

Sartorius Factory Series - Page 116

Sartorius Factory Series
213 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
116
Practical Example
Checkweighing samples of 170 g, with an allowable tolerance of 5 g and +10 g. Printout of upper and lower
tolerance limits. Weighed values are printed out automatically when stability is reached and the weight value is within
the checkweighing range.
Settings (changes in the factory settings required for this example):
Setup: App(lication parameters): Application 2: Checkweighing: Automatic printout of OK values: On
Step Press key(s) Display/Output
(or follow instructions)
1. Turn on the scale and configure e
the settings as indicated above
2. Delete previous values, if necessary c
3. Prepare a container Place empty container
for the samples on the scale
4. Tare the scale t
5. Enter initialization values
Param. soft key
6. Accept target value Place ideal sample
(in this example: 170 g) in container
7. Save target and unload l soft key
the scale Remove ideal sample
from the scale
8. Enter value for lower limit 1 6 5
(170 g 5 g) and save l soft key

Table of Contents

Other manuals for Sartorius Factory Series

Related product manuals