91
Operating the Scale
Practical Example
Determining Animal Weight With Automatic Start of 20 Subweighing Operations for Averaging; Automatic Printout of the
Number of Subweighing Operations and of the Animal Weight
Settings (changes in the factory settings required for this example):
Setup: App(lication parameters): Application 1: Animal weighing: Animal activity: Active
Setup: App(lication parameters): Application 1: Animal weighing: Decimal places in calculated result display: 2 decimal places
Setup: App(lication parameters): Application 1: Animal weighing: Printout: Average and calculated values
Setup: Printout: Application-defined output: Auto print upon initialization: All values
Step Key (or instruction) Display/Output
1. Delete previous setting if necessary c
2. Prepare a container (cage) Place empty cage
on the scale
3. Tare the scale t
4. Enter number of subweighing 2 0
operations for averaging
5. Save number
mDef soft key
6. Weigh the first animal Place 1st animal in cage weight value fluctuates due to animal activity
7. Start automatic animal weighing Start soft key
The scale delays starting the When this criterion is met, the
subweighing operation until subweighing series begins
three successive subweights lie
within the range defined for
an “active” animal