76 Troubleshooting
Err no.: Type: Usertext:
9108 E The ambient pressure measured is out of tolerance
9109 E The V2 valve for the reference tank is not operating properly or the reference tank is full
9110 E The internal reference tank is full
9112 E The test was subsequently cancelled because the operating pressure after the measurement was below 4 bar (58.02 psi)
9113 E The volume of the valve unit is 0 ml
9114 E During the pressure drop test, the pressure rose by more than 2 mbar (0.03 psi) - The temperature increased.
9115 E During the pressure drop test, the pressure drop exceeded the limit - The system leaks
9116 E The pressure drop test failed because the pressure drop exceeded the limit
9117 E Internal error
9118 E During calibration of the internal pressure sensor, the operating pressure for the selected pressure range is insufficient -
Increase the operating pressure
9119 E During calibration of the internal pressure sensor, all pressure values measured were the same so it is not possible to cali-
brate
9120 E During calibration/adjustment of the internal pressure sensor, the difference between two reference pressures is too small
for calibration to be performed.
9121 E During pressure buildup, the nominal pressure was exceeded
9122 E Timeout for pressure buildup exceeds the time limit
9123 E Internal error
9124 E During the stabilization time the system volume could not be stabilized sufficiently
9125 E The pressure drop at the end of stabilization exceeds the limit
9126 E Error while venting Warning: The system is probably still pressurized!
9127 E The loss of pressure during volume measurement exceeds the limit - Please check system for leakage - Please check config-
uration
9128 E The net volume determined is below 5 ml - Start cleaning cycle
9129 E During volume measurement (with an external reference tank > 6000 ml), the net volume measured was so large that it
was not possible to execute the test
9130 ...
9131
E During volume measurement (without an external reference tank) the net volume measured was so large that it was not
possible to execute the test
9132 E The volume measurement yielded a net volume greater than 12 l for the bubble point test or 50 l for the complete test
9133 E A considerable loss of pressure occurred at the beginning of the diffusion test - No filter housing was connected - The filter
cartridge is missing
9134 E During the diffusion test, a pressure increase occurred that may be caused by an increase in the temperature -Please wait
for the temperature to equalize -Check the temperature of the wetting liquid
9135 E During the diffusion test, the diffusion exceeded the limit -Please check the system for leakage -Check the configuration
-Rewet the filter
9136 E The last pressure value during the diffusion test exceeded the nominal pressure -Temperature increased during measure-
ment
9137 ...
9138
E During the bubble point or multi-point diffusion test, 5 consecutive pressure values exceeded the nominal pressure by 500
mbar (7.25 psi) -Reduce the operating pressure
9139 E The diffusion at the first bubble point pressure increment exceeds the limit - Check the system for leakage - Enter the cor-
rect test class - Adjust the diffusion limit
9140 E The diffusion at the second bubble point pressure increment exceeds the limit - Check the system for leakage - Enter the
correct test class - Adjust the diffusion limit
9141 E The diffusion from the third or higher bubble point pressure increment exceeds the limit - Therefore, the bubble point was
not recognized - Check the system for leakage - Enter the correct test class - Adjust the diffusion limit
9142 E Pressurization of housing has failed - Increase the operating pressure
9143 E The internal power supply is partially defective
9144 E The time it takes for stabilization exceeds the limit
9145 E During calibration of the temperature sensor, the commands were sent in an invalid sequence
9146 E During calibration of the temperature sensor, the type of sensor has been changed in between
9147 E During calibration of the temperature sensor, two identical temperatures were measured
9148 E During the bubble point test, the effective measuring time for the pressure curve was 0