Troubleshooting 75
Err no.: Type: Usertext:
6115 E Not possible to start visualization, as no test is running
6116 E Automatic program selection not possible as no program or report field input available
6117 W -Cancel ‘Error’ and ‘Ready’ states; -Cancel ‘Drying, internal/external adjustment not performed’; -Cancel ‘Test repeat, PLC
programs, last result’; -Abort program in progress Do you want this to be carried out?
6118 W Once you start the following procedure, it must be performed completely. Do you want to start this procedure?
6119 E Password was entered incorrectly too often. Please inform your system administrator.
6120 E New user needs to enter new password.
6121 E Temperature informations: <ARG1> <ARG2>
6122 E <ARG1> ----------------------------------------------------- Temperature informations: <ARG2> <ARG3>
6123 E No test result.
6124 E Cleaning and drying record not available.
6125 E Your password does not have the required minimum length of <ARG1> character(s); please enter new password.
6126 R The drying process was completed. - Close shutoff valve (f) - Disconnect “Venting 1” tubing from pressure tank (g) - Dis-
connect remaining tubing from pressure tank (g). Caution: Tank (g) is being vented - Please disconnect all tubing from the
connectors b-e.
6127 E The user file of the management unit is corrupt.
6901 ...
6902
E Printer server program cannot be addressed.
6903 E Unknown print job.
6904 E Job cannot be sent to printer server.
6905 E Response cannot be recieved from printer server.
6906 E Invalid print data.
6907 E Print job list is full.
6908 E Invalid test number for printing
6909 E Cannot write print file.
6910 E Cannot read print file.
6911 E Insufficient main memory for sending print job
6912 E Unknown error occurred during printing.
6999 E Internal error! <ARG1>
8000 ...
8300
E Communication error with test unit
9001 E It is not possible to start the selected procedure at the moment, because the former procedure has not yet been completed
- Please try again later
9004 E Only the internal reference tank is active
9005 E Both the internal and external reference tanks are active
9006 E The basic configuration data were re-initialized: (Address of the test unit, serial number, software version of the test unit) -
when the unit was initially operated at the factory - when the format of the data mentioned above was changed
9010 ...
9011
E The raw data of the graphs were not sent to the processing unit because the required component currently does not exist
9012 E The operating pressure ist too low. Please increase the operating pressure.
9013 E The operating pressure ist too high. Please reduce the operating pressure.
9014 E The operating pressure is not sufficient. Please increase the operating pressure.
9015 E The operating pressure is not stable. Please wait.
9101 E Error while venting Warning: The system is probably still pressurized!
9102 E The required test pressure exceeds the calibration range of the pressure sensor - Recalibrate using a wider calibration range
(0-8 bar/0-116.03 psi) - Work with a lower test pressure
9103 E The +12 V power supply is out of tolerance.
9104 E The -12V power supply is out of tolerance
9105 E The +24 V power supply is out of tolerance
9106 E Valve V1 does not close correctly.
9107 E No compressed air supply is available or valve V1 does not open correctly