Technical Data
13 Technical Data
13.1 Dimensions and Weight
Unit Value
Dimensions (L x W x H) mm 241.3 x 139.9 x 177.4
Weight kg 2.2
13.2 Ambient Conditions at the Installation Site
Unit Value
Conventional laboratory rooms, max. height above sea level m 2000
Temperature
Operation °C +5 – +40
Storage °C -40 – +60
Relative humidity, non-condensing
Operation % 25 – 85
Storage % 10 – 90
13.3 Power Supply
13.3.1 Device
Unit Value
Voltage V
DC
24
Current consumption A 2.5
Power supply only by Sartorius power supply unit WDS060240
13.3.2 Power Supply Unit
Unit Value
Type
Sartorius power supply unit only, model WDS060240
Primary
Voltage V
AC
100 – 240
Frequency Hz 50 – 60
Current consumption, maximum A 1.6
Secondary
Voltage V
DC
24
Current, maximum A 2.5