102 Scantron SelfScore Scanners
Procedure Overview
To recalibrate, perform the following:
• Prepare for recalibration
• Navigate the background menu
• Perform a Top Head Calibrate function
• Perform a Bot Head Calibrate function (if your
scanner is equipped with a bottom read head)
• Exit the background menu
Instructions for the above tasks are found in the pages
that follow.
Preparing for Recalibration
For best results, perform the following before
calibrating:
• Clean the scanner and the read head(s) as outlined
in Chapter 7, “Chapter 7: Cleaning the Scanner”.
• Ensure that the calibration forms are clean and
undamaged.