4 •
Administration of user data Access data for users of the device is administered in the seca 115 PC
software supplied. A user PIN for the seca 515/514 is generated automatically
when user accounts for the seca 115 are created. The administrator also has
the option of specifying a user PIN him or herself.
The device can only be configured with administrator rights. An initial adminis-
trator PIN for the device is provided. It can only be changed on the device.
The creation and administration of user data is only necessary if the seca
patient database of the
seca 115 PC software is to be accessed from the
device.
Data transmission and network
functions
The device is networkable. The network connection allows the device to use
both the seca patient database and the special print function of the seca 115
PC software.
The special print function of the
seca 115 PC software makes it possible to
start printing out a detailed results report directly on the medical Body
Composition Analyzer
seca 515/514.
Alternatively to the Ethernet link, seca mBCAs and the
seca 115 PC software
can communicate wirelessly via seca 360° technology. For this purpose, the
seca 360° wireless USB adapter 456 (included in the scope of delivery) must
be connected to a PC on which at least the application software of the
seca 115 is installed.
seca 360° stadiometers can wirelessly transmit measured results to the
device.
The device has the following interfaces:
• on the weighing platform
– network connection (Ethernet)
• on the touchscreen display
– internal seca wireless module
– USB interface for connecting a USB memory stick (included in the
scope of delivery)
Compatibility This device (software version 1.1 from Build 550) is compatible only with ver-
sion 1.4 from Build 560 of the
seca 115 PC software. There is no downward
compatibility with older versions of the seca 115. For a summary of technical
modifications, see the section entitled “Technical modifications” on page 56.
seca mBCAs with older device software can be updated. Go to
www.seca.com and contact your local seca service partner.
1.3 User qualification
Administration/network connection The device may only be set up and incorporated in a network by experienced
administrators or hospital technicians.
Measuring mode The device and the seca 115 PC software may only be operated by persons
with sufficient specialist expertise.