7.5 Print
8. OPTIONAL ACCESSORIES
Problem Cause Remedy
Print function not available
Software versions incompatible
Administrator: use compatible software
version:
device min. software version 1.1
PC software min. software version 1.4
No network connection set up between
device and
seca 115 PC software
•Set up
seca 360° wireless network
(“Setting up the seca 360° wireless
network” from page 40)
Set up Ethernet connection (“Integrate
device in an Ethernet network” from
page 37)
No seca CLS network set up
Administrator:
set up seca CLS network (see “Setting up
seca CLS network (via Ethernet only)” on
page 38)
A
seca 115 PC software service (medi-
ator service, calculation service, image
service, print service) has not started
Administrator:
start seca 115 PC software service
manually
Results report not being
printed
PC printer not switched on Switch on PC printer
The PC on which the
seca 115 PC
software is installed has not been
switched on
Switch on PC
No link set up between the
seca 115
PC software and the PC printer
Set up link between
seca 115 PC software
and the PC printer
No personal data in results
report
In the case of repeat measurement: no
seca patient file assigned
Call up the
patient tab and assign seca
patient file (see Instructions for Use for
Physicians and Assistants)
On initial measurement: seca patient
file not created yet
Call up the
patient tab and create seca
patient file (see Instructions for Use for
Physicians and Assistants)
No bioimpedance parameters
in results report
No bioimpedance measurement
performed
Call up the
bia tab and perform the mea-
surement (see Instructions for Use for
Physicians and Assistants)
No height in results report
No height entered on device
Call up the
weight/height tab and enter
height (see Instructions for Use for
Physicians and Assistants)
No height transmitted by the
seca 360°
stadiometer
Measure the patient's height again and
press the
send/print button (see Instruc-
tions for Use for Physicians and Assistants)
Accessory Article number
Measuring stations
seca 285
seca 284
Length measuring rods
seca 274
seca 264
Country-specific versions
Country-specific versions
Country-specific versions
Country-specific versions
seca analytics 115 PC software Application-specific license packages
seca 360° wireless USB adapter 456 USB wireless adapter 456-00-00-009
Crossover adapter for Ethernet cable 08-06-16-469-119