Portable X-ray Units
Service Manual
SM-1076R2
30
Table 3-3
Filament Current Manual Calibration Data
mA Stations 40 kVp 80 kVp 110 kVp (or 100 kVp)
5mA
6.4 mA
M
A
L
L
8mA
M
A
L
L
FOCAL
O
T
10 mA *
SPOT
12.5 mA * *
16 mA * *
20 mA * *
5mA
6.4 mA
8mA
10 mA
12.5 mA
16 mA
LARGE
F
O
C
A
L
20 mA
FOCAL
SPOT
25 mA
O
T
32 mA
40 mA *
50 mA *
64 mA * *
80 mA * *
100 mA * *
NOTES: 1) Some of the values with * should be entered manually following the numeric trend, the values with -- are not calibrated.
2) mA and kVp ranges depend on the Portable Unit model.
2.0 kW Units are calibrated at 40, 80 and 100 kVp.
4.0 kW and 8.0 kW Units are calibrated at 40, 80 and 110 kVp.