n Setting the function of the EVENT2 output terminalSELECT EVENT2
Not used in CL model.
Do not change the factory default value.
Sets the control mode of the [EVENT2]
output (relay 2).
Input the value with the numeric keypad
and press [Enter].
Set Value Function
0 The EVENT2 output is controlled with the [EVENT2] set value.
1 The EVENT2 output is used as an error output signal.
If the [SELECT EVENT2] function is used, the corresponding [EVENT2] parameter
is disabled.
n Setting the operation mode of the coolerCOOLER MODE
Select a cooler operation mode from the
three modes below. (Initial value is 2:
Auto)
Input the value with the numeric keypad
and press [Enter].
Set Value
Cooler Function
0 OFF Force OFF
1 ON Force ON
2 AUTO Automatic operation
Selecting the automatic operation causes the cooler to be turned on or off
automatically when the difference between the temperature setting and the
room temperature is between 10 °C and 14 °C. The temperature inside the oven
may temporarily show uneven dispersion when the cooler is turned on or off.
Select "Force ON" for a high-sensitive analysis that can be affected by this
temporary change.
n Setting the temperature ready checkREADY CHECK
Sets whether to wait for displaying
[READY] until the oven temperature
stabilizes at the set temperature or not.
26
CTO-40C CL
3 Operation