3
To confirm, input the same password
that was input in the step 2 again.
When the input is completed and the
password is changed, "PASSWORD
CHANGED" is displayed.
If the password input is not correct,
"PASSWORD WRONG" is displayed. At
this stage, the password has not been
changed yet.
4
Press [Enter].
The initial screen is displayed.
n
Setting the operation level of leak sensor《LEAK THOLD》
This function is not used in the CL model.
Recalibration operation is executed by service personnel.
Sets the threshold value of the leak sensor.
Input the value with the numeric keypad
and press [Enter].
Adjust it when a leak sensor check results
in a fail.
n Calibrating the gas sensor《GAS SENS CALIB》
This function is not used in the CL model.
Recalibration operation is executed by service personnel.
Runs the gas sensor calibration.
Adjust it when a gas sensor check results
in a fail.
40
CTO-40C CL
3 Operation