6.2 Error Message Shown in the "Status Window"
6-5
IRAffinity-1
There are more error messages, but in normal use they seldom appear. If undefined errors in the table above
frequently occur, the instrument may have a software or hardware problem. Please contact the Shimadzu
branch.
Error Message Meanings and Operation
Laser offline Laser inoperative. Please contact your SHIMADZU
representative.
Laser Failure If the laser indicator of status window shows red, laser is
damaged. Call Shimadzu branch.
No beamsplitter Beamsplitter is not mounted or not mounted correctly.
Please check.
Lamp offline. Please change it. If the lamp indicator of status window shows red, selected
lamp is disconnected. Call Shimadzu branch.
Timeout on initialization During initialization, communication between interferometer
and PC is not active. Check if interferometer power is not ON
or USB cable is disconnected.
Instrument Timeout Error Communication between interferometer and PC is not active.
Check if interferometer power is not ON or USB cable is
disconnected.
No valid range object Wavenumber range set file ("*.rng") may be damaged.
Please delete the ".rng" file and make a new range set file.
N2 information in EEPROM is invalid Information for the liquid nitrogen on EEPROM was
abnormal. Please contact your SHIMADZU representative.
Evaluation Scan Error: Internal counter
exceeded
Evaluation of data detected abnormality.
External trigger is not detected.
Measurement is aborted.
Outside trigger signal could not be detected.
Measured data intensity is too high (×16) Intensity of the data was too high. Gain setting was too high
or excessive noise was detected.
Measured data intensity is too low The light at the detector was too weak.