3 Starting and Stopping the Instrument
3-3
3.2 Starting the instrument (Daily use)
The routine procedure to start the instrument is described below.
1 Turn on the SOURCE switch on the switch panel.
2 Confirm that the temperature control indication on the switch panel is 40C.
3 Confirm that the vacuum pump is operating.
4 Check the water level in the argon gas bubbler.
(See "4.1 Argon gas bubbler".)
5 Adjust the argon gas flow rate to 1L / min. using the BYPASS valve.
(See "4.2 Checking gas flow rate".)
6 Turn on the printer.
7 Turn on the monitor.
8 Carry out recalibration.
(See "4.7 Recalibration".)
9 Carry out master curve recalibration when necessary.
(See "4.8 Master curve recalibration".)
10 Carry out check analysis using the sample for checking.
(See "4.9 Check analysis".)