EasyManuals Logo

Shimadzu PDA-7000 User Manual

Shimadzu PDA-7000
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
4 Operation
4-11
4.7 Recalibration
To calculate the recalibration coefficient, carry out the recalibration analysis.
1 Select Analysis - Recalibration on the menu bar.
2 Input the name of the analytical group in the Group textbox or select it by
pressing S4:Group key.
3 Select the name of the recalibration sample to be analyzed in the Sample combo
box.
Screen 4.3 Recalibration
4 Set the sample on the spark stand and hold it by the sample holder.
5 Start the discharge by pressing 1:Start key or the start button on the spark stand.
6 Take out the sample, and change the position of the sample to be analyzed.
Then, repeat procedures 3 and 4 specified times.
7 Repeat procedures 3 to 6 until the analyses of all the recalibration samples are
completed.
8 Press 3:Cal. Key.
The recalibration coefficients (α, β and k values) are calculated and they are
automatically stored in the recalibration information.
In the case of one point recalibration, coefficients α and β are not initialized. In
the case of two-point recalibration, k is initialized to 1.0.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu PDA-7000 and is the answer not in the manual?

Shimadzu PDA-7000 Specifications

General IconGeneral
BrandShimadzu
ModelPDA-7000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals