5 Maintenance
5-30
5.10 Waste discharge
After the maintenance of the instrument, carry out waste discharge to stabilize
emission.
1 Select Prepare - Waste Discharge on the menu bar.
Screen 5.3 Waste discharge
2 Set the sample on the spark stand and hold it by the sample holder.
3 Start the discharge by pressing 1:Start or the start button on the spark stand.
4 After repeating the discharge specified times, Content Analysis screen is
displayed automatically.