4 Operation
4-14
4.9 Check analysis
Confirm whether the content percentages of the elements are correctly measured,
using a sample with known composition.
1 Select Analysis - Cont. on the menu bar.
2 Input the name of the analytical group in the Group textbox or select it by
pressing S4:Group key.
Screen 4.5 Content analysis
3 Set the sample on the spark stand and hold it by the sample holder.
4 Start the discharge by pressing 1:Start key or the start button on the spark stand.
5 Take out the sample, and change the position of the sample to be analyzed.
Then, repeat procedures 3 and 4 specified times.
6 Confirm that the correct content percentages are obtained.