3 Starting and Stopping the Instrument
3-5
12 Carry out master curve recalibration when necessary.
(See "4.8 Master curve recalibration".)
13 Carry out check analysis using the sample for checking.
(See "4.9 Check analysis".)
3.4 Shutting down the instrument (Daily use)
The routine procedure to shut down the instrument is described below.
1 Turn off the SOURCE switch on the switch panel. Keep the MAIN, AIRCON, and
CONSOLE switches ON.
2 Place an analysis sample for blank analysis on the sample plate of the spark
stand and prevent air or debris from entering into the emission chamber.
3 Adjust the BYPASS valve to reduce the argon gas flow rate.
(See "4.2 Checking gas flow rate".)
4 Turn off the printer.
5 Turn off the monitor.