5 Maintenance
5-9
5 Wipe off the grease on the stand base
surface with gauze.
6 Remove evaporant in the dust cup
using a vacuum cleaner.
Before using the vacuum cleaner to clean the emission chamber, be
sure to remove the vinyl hose.
If the vacuum cleaner is used without removing the hose, water in the
gas bubbler may rise into the stand, and thus could damage the
instrument.
7 Detach the dust cup, remove evaporant under the cup using a vacuum cleaner and
dry-wipe the area.
The dust cup is made of ceramics, so dropping it or hitting it against
hard objects can damage it.
Also, if a damaged dust cup is used it can cause the analysis results
to be abnormal, so under no circumstances should a damaged dust
cup be used.