MOBILETT Plus HP SPR8-220.061.03 Page 28 of 28 Siemens AG
Rev. 03 09.03 CS PS 24 Medical Solutions
10 - 28 Mechanical systems
Install the new friction linings according to (Fig. 31 or Fig. 32). There are two different
vertical friction linings. Both are equally suitable for the unit.
Reinstall the upper counter bearing.
Tighten the screws.
Check that there is no braking action on the counter bearings.
Tighten the adjustment screw. Make sure that the spring mechanism is under tension
before releasing the ceiling support.
Attach the column cover and side covers and the cover with the cassette compartment.
Check parallelism (see section "Parallelism").
Adjust counterbalance (see section "Counterbalance").
Adjust friction linings (see section "Adjusting the friction linings").
Fig. 31 Fig. 32