MOBILETT Plus HP SPR8-220.061.03 Page 28 of 28 Siemens AG
Rev. 03 09.03 CS PS 24 Medical Solutions
10 - 28 Mechanical systems
• Install the new friction linings according to (Fig. 31 or Fig. 32). There are two different
vertical friction linings. Both are equally suitable for the unit.
• Reinstall the upper counter bearing.
• Tighten the screws.
• Check that there is no braking action on the counter bearings.
• Tighten the adjustment screw. Make sure that the spring mechanism is under tension
before releasing the ceiling support.
• Attach the column cover and side covers and the cover with the cassette compartment.
• Check parallelism (see section "Parallelism").
• Adjust counterbalance (see section "Counterbalance").
• Adjust friction linings (see section "Adjusting the friction linings").
Fig. 31 Fig. 32