Installation of chromatograph
INSTALLATION LOCATION
You must observe the following information
under Technical data.
Permissible ambient temperature
Permissible relative humidity
Power supply, power consumption
Lightning protection
Gas connections, gas purity
Purging air, control air
Space requirements
600 mm on both sides of chromatograph from
side wall or next chromatograph
Distance from wall at rear at least 100 mm
Distance from ceiling at least 200 mm
Swivelling radius of door 600 mm.
There are three manners in which the PGC
302
edition II can be mounted:
M8 threaded holes in the rear panel.
Caution:
The max. thread depth is
10 mm!
Mounting rails (option)
Rack (option)
536
1056
0
420
4 x M8
640
1014
0
420
80
10
Installation of chromatograph
2
3INSTALL.CHP, Stand Dezember 22, 1999