Correction of errors
in Register 5, V5.0
Register 5
Concerns All chromatographs
Chapter Detectors
Section Baseline monitoring
Paragraph Third
This should read:
... Enter the zero limits as an attenuation
(-21 to 21) which must not be fallen below or
exceeded.
The subsequent note must then read as follows:
... If you do not require this monitoring func-
tion, enter -21 for the lower zero limit and 21
for the upper zero limit.
Register 5
Concerns All chromatographs
Chapter Integration
Section Two chromatograms
simultaneously on screen
Paragraph
This paragraph should start with:
Whilst the chromatograph is displaying the list
of stored chromatograms, press the key
[
F9: Two chromatograms]. Select the first
chromatogram using ...
Register 5
Concerns All chromatographs
Chapter Integration
Section Enlarging a section (zoom)
Paragraph Third (Abort)
This should read:
You can abort marking of the section at any
time by pressing the key [
ESCAPE] (right
mouse button).
Register 5
Concerns All chromatographs
Chapter Appendix A:
All BASIC commands
Section Command IF...THEN...ELSE
Paragraph First
This should read:
20 I F C<20 THEN DO(1)=1 ELSE DO(1)=0
3 0 C = C + 1
Register 5
Concerns RGC 202,
process and laboratory versions
Chapter Operation from where
Section Complete page
Paragraph
We do not use the interface designations A to E any
more! Furthermore, switching over of interface B from
TTY to V.24 via interface C no longer exists.
Please read Register 3, Connection of control unit and
possibly the following pages for the current arrange-
ment.
Register 5
Concerns RGC 202, Laboratory version
Chapter Appendix C,
Parameterization of autosampler
Section Position of solvent vial in belt
Paragraph Last sentence
This should read:
The solvent vial (purging solution) must al-
ways be inserted into position S (at the second
position A with the AS 32).