Register 5
Concerns RGC 202, process version
Chapter Updating of retention window
Section Viewing corrected retention times
Paragraph
The key F7 must read:
[
F7: correc. Ret.Time] instead of
[
F7: currrent Ret.Time]
The corrected retention times can also be reset by
pressing [
F8: Reset ret.time
]. The retention win-
dows which you have entered in the component list are
then effective again.
Register 5
Concerns RGC 202, process version
Chapter Appendix A:
All BASIC commands
Section Command
FOR...TO...STEP...NEXT
Paragraph Sixth (second example)
This should read:
10 FOR I =5 TO 2 STEP -1
Register 5
Concerns RGC 202, process version
Chapter Appendix A:
All BASIC commands
Section Command ALA/ALB
Paragraph Flag table
The flag table has been extended:
Flag table
Flag Function Flag Function
1 Controller 1 ready 33 Temp.prog. 4 running
2 Controller 1 alarm 34 Temp.prog. 5 running
3 Controller 2 ready 35 Temp.prog. 6 running
4 Controller 2 alarm 36 Temp.prog. 7 running
5 Controller 3 ready 37 Temp.prog. 8 running
6 Controller 3 alarm 38 Temp.prog. 9...11 running
7 Controller 4 ready 39 Controller 9...11 ready
8 Controller 4 alarm 40 Controller 9...11 alarm
9 Controller 5 ready 41 BASIC alarm 1
10 Controller 5 alarm 42 BASIC alarm 2
11 Controller 6 ready 43 BASIC alarm 3
12 Controller 6 alarm 44 BASIC alarm 4
13 Controller 7 ready 45 BASIC alarm 5
14 Controller 7 alarm 46 BASIC alarm 6
15 Controller 8 ready 47 BASIC alarm 7
16 Controller 8 alarm 48 BASIC alarm 8
17 Simultaneous run 49 Detector alarm
18 Maintenance operation 50 Upper zero limit
19 Periphery faulty 51 Lower zero limit
20 Analog data output 52 Result ready
21 Gas supply 53 Analysis not running
22 Interface 54 External alarm
23 Internal temperature 55 Calibration running
24 5-V current 56 Calibration within cycle
25 24-V current 57 Calibration, last cycle
26 Memory overflow 58 Calibration, absolute
27 Leak detector 59 Updating of retention time
28 Purging pressure 60 Time events not running
29 Process operation
30 Temp.prog. 1 running
31 Temp.prog. 2 running
32 Temp.prog. 3 running
HPLC
30 Pump A alarm
31 Pump B alarm
32 Pump C alarm
33 Eluent A empty
34 Eluent B empty
35 Eluent C empty
36 Pulsation A alarm
37 Pulsation B alarm
38 Pulsation C alarm
Miscellaneous
5DIVERSE.CHP, Stand Dezember 22, 1999
29