Technical data
TECHNICAL DATA
Climatic conditions
Permissible ambient temperature
5 to 40 °C
Upward or downward deviations
possible depending on design
Permissible storage temperature
- 30 to 70 °C
Permissible relative humidity without purging
With FID, FPD, ECD
Max. 80 % at 25 °C
Max. 60 % at 40 °C
With TCD
Max. 90 %
Protection against dust and humidity
to EN 60529 / VDE 0470, part 1 / IEC 529
IP 54
Power supply
Power supply
230 or 115 V +10 % -15 %, 48 to 63 Hz
Power consumption
Max. 2 kVA in heating-up phase
Approx. 0.3 / 0.5 kVA with 2 ovens and 100 / 180 °C
External mains fuse, for 230 / 115 V
16 / 20 A
Dimensions and weights
Dimensions in mm (w x h x d)
600 x 1200 x 420
With FDV at side
790 x 1200 x 420
Weight
Approx. 90 kg
Electromagnetic compatibility
Radiointerference suppression
To VDE 0875 / EN 55011 / CISPR 11 limit class B
EMC
To DIN VDE 0843 / IEC 801, Part x
Conducted interferences on AC power supply lines
To Part 4 (burst) 2 kV
To Part 5 (µs pulses), line against line 1 kV
To Part 5 (µs pulses), line against earth 2 kV
Conducted interferences on signal lines
To Part 4 (burst) 1 kV
Interference immunity to discharges of static electricity
To Part 2 (ESD) 8 kV
Interference immunity to fields
To Part 3 and Part 6 10 V/m
Safety
Electric safety
DIN VDE 0411 / IEC 1010
Explosion protection
EEx pde IIB+H
2
, T2, T3 or T4,
division 1 or 2
Ovens
Number
1 or 2
Air circulation
None
N
2
purging
Possible
Dimensions (w x h x d)
220 x 260 x 195 (or 145) mm
3
Max. heating capacity
800 VA
Temperature range
60 to 260 °C (*1)
with cooling unit
5 to 180 °C (*1)
Max. temperature difference between ovens
100 K
Temperature stability
± 0.1 K (60 to 180 °C)
Temperature accuracy
± 3 K (60 to 180 °C)
Temperature gradient
± 2 K (at 180 °C)
Variation in retention time per
10 K change in ambient temperature
Approx. -0.3 %
Heating-up time from 30 to 160 °C
60 minutes
Columns and gases
Column type
Packed columns with 3 or 4 mm (
1
8
") diam.
Capillary columns with 0.20 to 0.53 mm internal diam.
Column switching
Multi-dimensional chromatography
with backflushing and live cutting
for packed and capillary columns,
or with valve switching for packed columns
Gas connections
Pipe 3 x 0.5 (approx. 1/8")
except purging gas, 6 mm (
1
4
") Swagelok
liquid injection valve, 3 mm (1/8") Swagelok
and heated gas connections 6 mm (
1
4
") Swagelok
Gas purity (minimum requirements)
99.999 %
Control air and purging air
Type of gas
Compressed air, free of oil, water and dust
Permissible relative humidity
Dew point more than 5 K
below the lowest ambient temperature
Control air and purging air for Division 1
4 to 6 bar
Control air at least 2 bar above carrier gas pressure
Purging air for Division 2
restricted to 10 ... 200 l/min
depending on purging situation
Corrosion protection purging
Approx. 2 to 10 l/min
Housing volume of pressure enclosure
Approx. 140 l
Enclosure prepurging
130 to 200 l/min for 12 minutes
Operating phase
8 to 12 l/min
Sample
Solid particles
in liquid sample < 0,3 µm
in gaseous sample < 0,1 µm
(*1)
The following limit temperatures apply to chromatographs
with explosion protection according to the following tempera-
ture class
Max. 120 °C with T4
Max. 180 °C with T3
Max. 280 °C with T2
Two-stage monitoring by
Limit temperatures
Safety temperature limiter
Technical data
24
3DATEN.CHP, Stand Dezember 22, 1999