Optimization of time events program
for chromatogram printout
In a maximum of six time events program lines it is
possible to modify the measuring range, baseline offset
and chart drive.
These parameters only have an effect on the chromato-
gram printout. They have nothing to do with the par-
ameters for displaying the chromatogram on the re-
corder.
NOTE:
The time events program is a part of the method.
Select the function Raw data/Print in the selection dis-
play. Move the cursor to Print results or Print chroma-
togram for the desired channel. Press
.
Switchover time
The parameters in the line Start value become active
when printing is started. Switchover time is the respec-
tive time at which the new parameters are to apply.
Time 0 means that the complete line is ineffective.
Measuring range
The measuring range from 0 to 22 is the same as with
the recorder display.
Chart drive
The chart drive can be set in seven steps. Step 1 prints
approx. 30 minutes of chromatogram on one page, step
7 prints approx. 30 seconds.
The parameter Peak width in the display Peak recogni-
tion influences the print result. Steps appear in the
chromatogram if the peak width is selected too large.
The following applies:
Chart speed Max. bunching Minutes per page
1
2
3
4
5
6
7
60
30
20
10
5
2
1
30
15
10
5
2
1
0,5
IMPORTANT:
When printing automatically following each
run, the printout must be completed during
the run. Not all chromatograms are otherwise
printed out!
Baseline offset
The baseline offset from -99 to +99 % has the same
effect as with the recorder display. The percentage
value always refers to the full-scale value in the se-
lected measuring range.
F6: Printout parameters
Print chromatogram and results
5CHRODRU.CHP, Stand Dezember 22, 1999