Detectors
FID
Detection limit for
hydrocarbons referred to carbon
2 x 10
-12
g/s
Dynamic response
10
7
Automatic ignition
Max. operating temperature
280 °C (*1)
Combustion
and carrier gases
20 to 60 ml/min (H
2
, N
2
, He, Ar)
Combustion air
0.4 l/min
FPD
Detection limit for sulphur
2 x 10
-11
g/s
Detection limit for phosphorus
9 x 10
-13
g/s
Sulphur characteristic
Approx. quadratic
Linear range for phosphorus
10
5
Max. operating temperature, classified
180 °C (*1)
Max. operating temperature, non-classified
300 °C
Carrier gas
10 to 40 ml/min (N
2
, Ar)
Combustion gas
60 to 80 ml/min (H
2
)
Combustion air
100 to 150 ml/min
TCD
Detection limit
for ethane in H
2
at width at half peak height = 6 s
3 x 10
-9
g/ml (2 ppm)
Linear range
10
4
Cell volume, standard/micro
0.35 ml / 70 µl
Max. operating temperature, classified
180 °C (*1)
Max. operating temperature, non-classified
200 °C
Carrier gas
10 to 60 ml/min (H
2
, N
2
, He, Ar)
ECD
Detection limit for lindane
5 x 10
-14
g/s
Linearity
>10
3
Max. operating temperature, non-classified
350 °C
Activity (Ni 63)
370 MBq
Purging gas and carrier gas
(N
2
, 95 % argon/5 % methane)
Total 15 to 50 ml/min
Injection
Liquid injection valve
Injection volume
0.6 to 9.5 µl
Vaporization temperature
60 to 400 °C (*1)
Sample pressure
0.1 to 50 bar abs.
Diaphragm valve
Injection volume
0.2 to 2 ml
Max. operating temperature
160 °C
Sample pressure
0.9 bar to
2 bar abs. below control air pressure
Rotary slide valve
Injection volume
0.01 to 10 ml possible
Max. operating temperature
220 °C (*1)
Sample pressure
0 to 20 bar abs.
Electronics
CPU
Microprocessor
Intel 80C186-16
Clock
12.5 MHz
Dynamic RAM
512 Kbyte
Flash EPROM
431 Kbyte
Software
Loadable from control unit
Interfaces
Number
2 x TTY or RS 232
2 x RS 485 or RS 232
Transmission rate
Up to 38400 baud,
adjustable in software
Loadable drivers
RK512, D3964R, Modbus,
message printer, local operation
Networking
ChromLAN, redundant also possible
Control system coupling
TELEPERM M (RK512)
Honeywell (Modbus)
Inputs/outputs: basic configuration
Digital outputs 0.5 A
24, 16 of which for optional use
Digital inputs
16, 8 of which for optional use
Analog outputs
4 x 0/4 to 20 mA
750 max. load, 10-bit resolution
Vacant slots
5 x single width (single channel)
4 x single width (double channel)
Possible extensions with SIMATIC modules
32 digital outputs
24 V / 0.5 A
16 digital outputs
24 V / 2 A, double width
32 digital inputs
24 V
16 digital outputs
24 V / 0.5 A
and 16 digital inputs
24 V
8 analog outputs
0 to 20 mA or 4 to 20 mA
300 max. load, 10-bit resolution
8 analog inputs
0/4 to 20 mA, 12-bit resolution
Input resistance 25
Technical data
3DATEN.CHP, Stand Dezember 22, 1999