Page 186 SITRANS LUT400 (HART) – OPERATING INSTRUCTIONS A5E33329501
mmmmm
Parameters
To use TVT shaper via SIMATIC PDM:
a) Open the menu Device – Echo Profile Utilities and click on TVT Shaper. (For
more details see
TVT Shaper
in LUT400 Communications manual
1
.)
To use TVT shaper via local push buttons:
a) Navigate to Setup (2.) > Signal Processing (2.12.) > TVT Setup (2.12.3.) >
Shaper Mode (2.12.3.4.), and select ON.
b) From the TVT Setup menu, LEFT ARROW to the Signal Processing menu,
and DOWN ARROW to TVT Shaper. RIGHT ARROW to enter the TVT
Shaper menu and RIGHT ARROW to edit Breakpoint 1-8 (2.12.4.1.).
c) Open TVT Breakpoint 1 enter the TVT Offset value (between –50 and 50).
d) Go to the next TVT Breakpoint and repeat steps c) and d) till all desired
breakpoint values have been entered.
2.12.4.1. Breakpoint 1-8
2.12.4.2. Breakpoint 9-16
2.12.4.3. Breakpoint 17-24
2.12.4.4. Breakpoint 25-32
2.12.4.5. Breakpoint 33-40
1.
Communications for SITRANS LUT400 (HART) Manual (7ML19985NE01)
Values
Range: –50 to 50 dB
Default: 0 dB
Values
Range: –50 to 50 dB
Default: 0 dB
Values
Range: –50 to 50 dB
Default: 0 dB
Values
Range: –50 to 50 dB
Default: 0 dB
Values
Range: –50 to 50 dB
Default: 0 dB