Head
Hints
For image reconstruction of soft tissue, use kernel
H30s/H31s.
An automatic bone correction allows for improved
head image quality, without any additional postpro-
cessing.
In order to optimize image quality versus radiation
dose, scans are provided within a maximum scan
field of 300 mm with respect to the isocenter. No
recon job with a field of view exceeding those limits
will be possible. Therefore, patient positioning has to
be performed accurately to ensure a centered loca-
tion of the skull.
Contrast medium IV injection
Start delay 60 sec.
Flow rate 2 ml/sec.
Volume 50-60 ml