EasyManua.ls Logo

Siemens SOMATOM

Siemens SOMATOM
590 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Scan and Reconstruction
37
Neuro Modes
In addition to the standard collimations, the SOMA-
TOM Emotion 16-slice configuration provides a special
mode which is optimized for Neuro applications. Excel-
lent low contrast and detail resolution are achieved.
For spiral scans 4 x 0.6 acquisition mode is provided in
the range of the cerebrum. This approach shows a min-
imized partial volume effect, i.e. low level of artifacts
in the base of the skull or near vertebral bodies, as
0.6 mm detector rows are used and the narrow colli-
mation reduces scattered radiation.
One scan protocol is predefined for adults:
HeadNeuro using an acquisition 16 x 0.6 mm in the
base and an acquisition of 4 x 0.6 mm in the cere-
brum
We recommend using this special protocol for dedi-
cated Neuro examinations.
For fast standard examinations such as rule out of
hemorrhage or ischemia, the "Routine" protocol should
be used.

Table of Contents

Other manuals for Siemens SOMATOM

Related product manuals