574
Children
HeadAngio/
HeadAngio06s
Indications:
Spiral mode for head CT Angio studies, for example,
cerebral vascular abnormalities, tumors etc.
For SOMATOM Emotion 16-slice configuration:
A typical range of 6 cm covered in 3.70 sec.
For SOMATOM Emotion 6-slice configuration:
A typical range of 6 cm covered in 5.20 sec.
Emotion 16 HeadAngio 2
nd
reconstr.
kV 80
Effective mAs/
Quality ref. mAs
80
Rotation time 0.6 sec.
Acquisition 16 x 0.6 mm
Slice collimation 0.6 mm
Slice width 2.0 mm 1.0 mm
Feed/Rotation 14.4 mm
Pitch Factor 1.50
Increment 2.0 mm 0.7 mm
Kernel C30s C30s
CTDI
Vol
6.16 mGy
Effective dose Male: 0.20 mSv*
Female: 0.21 mSv*

Table of Contents