EasyManuals Logo

Siemens SOMATOM Manual&Nbsp;

Siemens SOMATOM
590 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #304 background imageLoading...
Page #304 background image
304
Spine
SpineThinSlice
Indications:
Spiral mode for thin slice thoracic and lumbar spine
studies, e.g. prolapse, degenerative changes, trauma,
tumors etc.
For SOMATOM Emotion 6-slice configuration:
A range of 10 cm will be covered in 32.30 sec.
F
o
r
S
O
M
A
T
Emotion 6 Spine
ThinSlice
2
nd
recon.
3
rd
recon.
kV 130
Effective mAs/
Quality ref. mAs
190
Rotation Time 1.0 sec.
Acquisition 6 x 1.0 mm
Slice collimation 1.0 mm
Slice width 3.0 mm 1.25 mm 1.25 mm
Feed/Rotation 3.3 mm
Pitch Factor 0.55
Increment 3.0 mm 0.8 mm 0.8 mm
Kernel B31s B31s B70s
CTDI
Vol
24.13 mGy
Effective dose Male: 3.64 mSv
Female: 7.23 mSv

Table of Contents

Other manuals for Siemens SOMATOM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SOMATOM and is the answer not in the manual?

Siemens SOMATOM Specifications

General IconGeneral
CategoryCT Scanner
Rotation Time0.25 seconds
Tube Voltage70-150 kV
Bore Size78 cm
Max Table Load250 kg
Gantry Aperture78 cm
Table Weight Capacity250 kg
X-ray Tube Voltage70-150 kV
Scan Time per Rotation0.25 seconds
SoftwareSyngo
ConnectivityDICOM

Related product manuals