Loading...
Specials
437
Biopsy
With this routine protocol images will be reconstructed
with three time 4.8/6.0 mm or one time 10 mm slice
thickness.
Emotion 16 Biopsy
kV 130
Effective mAs/
Quality ref. mAs
110
Rotation time 0.6 sec.
Acquisition 12 x 1.2 mm
Slice collimation 1.2 mm
Slice width 4.8 mm
Feed/Scan 0.0 mm
Kernel B31s
CTDI
vol
12.10 mGy
Emotion 6 Biopsy
kV 130
Effective mAs/
Quality ref. mAs
110
Rotation time 0.6 sec.
Acquisition 6 x 3.0 mm
Slice collimation 3.0 mm
Slice width 6.0 mm
Feed/Scan 0.0 mm
Kernel B31s
CTDI
vol
11.22 mGy