414
Vascular
WholeBodyAngio
Indications:
For CT Angio spiral studies of the whole body.
For SOMATOM Emotion 16-slice configuration:
A range of 80 cm will be done in 17.87 sec.
Emotion 16 WholeBody 2
nd
reconstr.
kV
110
Effective mAs/
Quality ref. mAs
90
Rotation Time 0.6 sec.
Acquisition 16 x 1.2 mm
Slice collimation 1.2 mm
Slice width 6.0 mm 1.5 mm
Feed/Rotation 28.8 mm
Pitch Factor 1.50
Increment 6.0 mm 1.0 mm
Kernel B31s B31s
CTDI
Vol
6.48 mGy
Effective dose Male: 9.33 mSv
Female: 10.11 mSv