EasyManua.ls Logo

Siemens SOMATOM

Siemens SOMATOM
590 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
376
Vascular
CarotidAngioVol
Indications:
CT angiography of the carotid arteries, e.g. carotid
stenosis or occlusions, coarse plaques abnormalities of
the carotids and vertebral arteries, etc.
Two recon jobs are predefined for reconstruction: the
first for axial, the second for coronal studies in 3D
images display view.
The coronal view images will be reconstructed as MIP
images.
For SOMATOM Emotion 16-slice configuration:
A range of 20 cm including the aorta arch will be cov-
ered in 9.53 sec.
For SOMATOM Emotion 6-slice configuration:
A range of 12 cm including the aorta arch will be cov-
ered in 12.27 sec.
Emotion 16 CarotidAngio 2
nd
reconstr.
kV 110
Effective mAs/
Quality ref. mAs
70
Rotation time 0.6 sec.
Acquisition 16 x 0.6 mm
Slice collimation 0.6 mm
Slice width 5.0 mm 3.0 mm
Feed/Rotation 14.4 mm
Pitch Factor 1.50
Increment 5.0 mm 3.0 mm
Kernel B31s B20s
CTDI
Vol
5.60 mGy
Effective dose Male: 1.28 mSv
Female: 1.41 mSv

Table of Contents