426
Specials
PolyTrauma/
PolyTrauma06s
Two combined ranges are predefined, head with neck
and thorax with abdomen.
For SOMATOM Emotion 16-slice configuration:
A scan range of 12/14 cm will be done in 13.36/9.68
sec.
Emotion 16 Head Neck
kV 130 130
Effective mAs/
Quality ref. mAs
240 150
Rotation time 1.0 sec. 1.0 sec
Acquisition 16 x 1.2 mm 16 x 1.2 mm
Slice collimation 1.2 mm 1.2 mm
Slice width 6.0 mm 5.0 mm
Feed/Rotation 10.6 mm 19.2 mm
Pitch Factor 0.55 1.00
Increment 6.0 mm 5.0 mm
Kernel H31s B50s
CTDI
Vol
57.60 mGy 16.80 mGy
Effective dose
Male:
Female:
2.63 mSv
2.88 mSv
3.08 mSv
3.37 mSv