504
Respiratory Gating
RespModBreathRate
Indications:
A spiral mode for patients with a respiration rate per
minute greater than 7.5, using respiration gating.
For SOMATOM Emotion 16-slice configuration:
A scan range of 23 cm will be covered in 94.15 sec.
For SOMATOM Emotion 6-slice configuration:
A scan range of 14 cm will be covered in 94.93 sec.
Emotion 16 RespMode 2
nd
reconstr.
kV 130
Effective mAs/
Quality ref. mAs
320
Rotation time 1.0 sec.
Acquisition 16 x 1.2 mm
Slice collimation 1.2 mm
Slice width 3.0 mm 3.0 mm
Feed/Rotation 2.5
Pitch Factor 0.13
Increment 2.1 mm 2.1mm
Kernel B31s B31s
CTDI
vol
35.84 mGy