198
Head
Orbit
Indications:
Spiral mode for orbital studies, e.g. fracture.
For SOMATOM Emotion 16-slice configuration:
A range of 50 mm will be covered in 8.51 sec.
For SOMATOM Emotion 6-slice configuration:
A range of 50 mm will be covered in 11.80 sec.
Emotion 16 Orbit 2
nd
reconstr.
kV 130
Effective mAs/
Quality ref. mAs
70
Rotation Time 1.0 sec.
Acquisition 16 x 0.6 mm
Slice collimation 0.6 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 7.7 mm
Pitch Factor 0.80
Increment 5.0 mm 0.7 mm
Kernel H70s H70s
CTDI
Vol
18.62 mGy
Effective dose Male: 0.50 mSv
Female: 0.54 mSv