Head
197
Hint
For low dose studies use only 20 mAs.
Emotion 6 Sinus 2
nd
recon.
3
d
recon.
kV 130
Effective mAs/
Quality ref. mAs
35
Rotation Time 1.0 sec.
Acquisition 6 x 1.0 mm
Slice collimation 1.0 mm
Slice width 5.0 mm 3.0 mm 3.0 mm
Feed/Rotation 5.1 mm
Pitch Factor 0.85
Increment 5.0 mm 3.0 mm 3.0 mm
Kernel H70s H30s H60s
CTDI
Vol
9.63 mGy
Effective dose Male: 0.20 mSv
Female: 0.25 mSv
Contrast medium IV injection
Start delay 45 sec.
Flow rate 2.0 ml/sec.
Total amount 75 ml