392
Vascular
ThorAngioECGSeq
Indications:
Sequential, ECG triggered study for location and extent
of aneurysms, dissection, and ruptures of the thoracic
aorta with typical scan range of 29.9 cm.
Hints
CARE Bolus may be used to optimize the bolus tim-
ing.
Set the ROI for monitoring scan in the aortic arch
with triggering threshold of 120 HU, or use manual
triggering.
Emotion 6 ThorAngioECG
kV 130
Effective mAs/
Quality ref. mAs
80
Rotation Time 0.8 sec.
Acquisition 6 x 1.0 mm
Slice collimation 1.0 mm
Slice width 1.0 mm
Feed/Scan 6.0 mm
Kernel B31s
Temp. resolution
*
*. depends on heart rate
up to 400 ms
CTDI
Vol
10.16 mGy
Effective dose Male: 6.92 mSv
Female: 9.00 mSv
Contrast medium IV injection
Start delay 10 – 30 sec.
Flow rate 2.5 ml/sec.
Total amount 80 ml