Configuration
Dose Display
The maximum displayed value of the Dose is set as a
default to 2000 mGy. It is possible to configure this
from 100-200 mGy under Options>Configura-
tion>Examination>Intervention. Exceeding this con-
figuration maximum value will not stop the scan, just
the Dose scale bar will change to red. If you start scan-
ning again the Dose scale bar will be reset.
Storage rate
All three CAREView images (Head, Center and Feet)
can be saved if you select it under Configura-
tion>Examination>Intervention prior to your exami-
nation.
The image storage rate is also configurable under
Option>Configuration>Examination>Intervention
depending on your system. You can choose between
an image storage rate of 1,2,4 (5 images/sec. for
SOMATOM Emotion 16/6-slice configuration ) and 1*3
images per second.