Page #436 background imageLoading...
Page #436 background image
436
Specials
Interventional CT - Biopsy
To facilitate CT interventional procedures, we created
dedicated multislice and single slice sequential modes.
Any of these protocols can be appended to a spiral pro-
tocol for CT interventional procedures, such as biopsy,
abscess drainage, pain therapy, minimum invasive
operations, joint studies, and arthrograms. Adjust the
mAs according to the body region before loading.
10 scans are predefined. You can repeat it by clicking
the chronicle with the right mouse button and select
repeat, or simply change the number of scans to 99
before you start the first scan.
You can “Append” any routine protocol after the inter-
ventional procedure for a final check and documenta-
tion, e.g. a short range of spiral scanning for the biopsy
region.
The table height can be adjusted to a minimum of
255 mm.
Zoom and pan of the images is possible within the
Biopsy Mode.

Table of Contents

Other manuals for Siemens SOMATOM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SOMATOM and is the answer not in the manual?

Siemens SOMATOM Specifications

General IconGeneral
CategoryCT Scanner
Rotation Time0.25 seconds
Tube Voltage70-150 kV
Bore Size78 cm
Max Table Load250 kg
Gantry Aperture78 cm
Table Weight Capacity250 kg
X-ray Tube Voltage70-150 kV
Scan Time per Rotation0.25 seconds
SoftwareSyngo
ConnectivityDICOM