EasyManuals Logo

Siemens SOMATOM Manual&Nbsp;

Siemens SOMATOM
590 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #305 background imageLoading...
Page #305 background image
Spine
305
SpineVol
Indications:
Spiral mode for thoracic and lumbar spine studies, e.g.
prolapse, degenerative changes, trauma, tumors etc.
Three recon jobs are predefined for reconstruction: the
first for soft tissue axial, the second for soft tissue sag-
ittal and third for coronal bone studies in 3D images
display view.
For SOMATOM Emotion 16-slice configuration:
A range of 15 cm will be covered in 26.04 sec.
For SOMATOM Emotion 6-slice configuration:
A range of 10 cm will be covered in 32.30 sec.
Emotion 16 Spine 2
nd
recon. 3
rd
recon.
kV 130
Effective mAs/
Quality ref. mAs
190
Rotation Time 1.0 sec.
Acquisition 16 x 0.6 mm
Slice collimation 0.6 mm
Slice width 3.0 mm 2.0 mm 2.0 mm
Feed/Rotation 6.2 mm
Pitch Factor 0.65
Increment 3.0 mm 2.0 mm 2.0 mm
Kernel B31s B20s B20s
CTDI
Vol
23.75 mGy
Effective dose Male: 4.42 mSv
Female: 4.70 mSv

Table of Contents

Other manuals for Siemens SOMATOM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SOMATOM and is the answer not in the manual?

Siemens SOMATOM Specifications

General IconGeneral
CategoryCT Scanner
Rotation Time0.25 seconds
Tube Voltage70-150 kV
Bore Size78 cm
Max Table Load250 kg
Gantry Aperture78 cm
Table Weight Capacity250 kg
X-ray Tube Voltage70-150 kV
Scan Time per Rotation0.25 seconds
SoftwareSyngo
ConnectivityDICOM

Related product manuals