442
Specials
CAREVisionBone
The images will be reconstructed and displayed in the
CARE View mode with three times 4.8/6.0 mm or one
time 10 mm slice thickness and a kernel of B50.
Emotion 16 CAREVision
kV 130
Effective mAs/
Quality ref. mAs
30
Rotation time 0.6 sec.
Acquisition 16 x 0.6 mm
Slice collimation 0.6 mm
Slice width 4.8 mm
Feed/Scan 0.0
Increment 1.0 mm
Kernel B50s
CTDI
vol
3.75 mGy
Emotion 6 CAREVision
kV 130
Effective mAs/
Quality ref. mAs
30
Rotation time 0.6 sec.
Acquisition 6 x 3.0 mm
Slice collimation 3.0 mm
Slice width 6.0 mm
Feed/Scan 0.0
Increment 1.0 mm
Kernel B50s
CTDI
vol
3.06 mGy