EasyManuals Logo

Siemens SOMATOM Manual&Nbsp;

Siemens SOMATOM
590 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #245 background imageLoading...
Page #245 background image
Thorax
245
Hints
For lung cancer evaluation, this protocol can be com-
bined with protocol “Neck Routine”.
Emotion 6 ThorFast 2
nd
recon.
kV 130
Effective mAs/
Quality ref. mAs
70
Rotation Time 0.8/0.6 sec.
Acquisition 6 x 3.0 mm
Slice collimation 3.0 mm
Slice width 6.0 mm 6.0 mm
Feed/Rotation 27.0 mm
Pitch Factor 1.50
Increment 6.0 mm 6.0 mm
Kernel B41s B70s
CTDI
Vol
7.14 mGy
Effective dose Male: 4.12 mSv
Female: 5.25 mSv
Contrast medium IV injection
Start delay 25 – 30 sec.
Flow rate 2.5 ml/sec.
Total amount 80 ml