Thorax
251
Hints
If you want to reconstruct thin slices every 10 or 20
mm instead of 15 mm as predefined, simply change
the Feed/Scan before loading the mode.
With studies of interstitial changes in the lungs, con-
trast medium is not necessary.
Emotion 6 ThorHRSeq
kV 130
Effective mAs/
Quality ref. mAs
100
Rotation time 1.0 sec.
Acquisition 1 x 1.0 mm
Slice collimation 1.0 mm
Slice width 1.0 mm
Feed/Scan 10.0 mm
Kernel B90s
CTDI
Vol
1.01 mGy
Effective dose Male: 0.49 mSv
Female: 0.63 mSv