Thorax
257
Hints
For lung cancer evaluation, this protocol can be com-
bined with protocol “Neck Routine”.
Low dose lung images are usually evaluated using
lung window setting. Soft tissue/bone window set-
tings may be used to detect the presence of calcifica-
tions in the nodules.
It is essential to use the same protocol for follow-up
studies to check for progression.
CARE Dose 4D is off as default because for LungLow-
Dose protocols the lowest mAs values are used.
Contrast medium IV injection
Start delay 30 sec.
Flow rate 2.5 ml/sec.
Total amount 50 – 70 ml