Abdomen
275
Emotion 6 AbdCombi 2
nd
recon.
kV 130
Effective mAs/
Quality ref. mAs
130
Rotation Time 0.8
Acquisition 6 x 1.0 mm
Slice collimation 1.0 mm
Slice width 5.0 mm 5.0 mm
Feed/Rotation 9.0 mm
Pitch Factor 1.50
Increment 5.0 mm 5.0 mm
Kernel B41s B30s
CTDI
Vol
16.51 mGy
Effective dose Male: 6.32 mSv
Female: 7.85 mSv
Contrast medium IV injection
Start delay 50 – 65 sec.
Flow rate 3.0 ml/sec.
Total amount 100 ml