Abdomen
283
For SOMATOM Emotion 6-slice configuration:
A range of 20 cm including liver, pancreas and kidney:
arterial phase acquired in 7.87 sec.
Emotion 6 Non Contrast
kV 110
Effective mAs/
Quality ref. mAs
75
Rotation Time 0.8/0.6 sec.
Acquisition 6 x 2.0 mm
Slice collimation 2.0 mm
Slice width 5.0 mm
Feed/Rotation 18.0 mm
Pitch Factor 1.50
Increment 5.0 mm
Kernel B41s
CTDI
Vol
5.33 mGy
Effective dose Male: 2.14 mSv
Female: 2.71 mSv